Method development for the determination of fluorine in water samples via the molecular absorption of strontium monofluoride formed in an electrothermal atomizer


Ozbek N., Akman S.

SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, cilt.69, ss.32-37, 2012 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 69
  • Basım Tarihi: 2012
  • Doi Numarası: 10.1016/j.sab.2012.03.003
  • Dergi Adı: SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.32-37
  • Anahtar Kelimeler: High-resolution continuum source electrothermal atomic absorption spectrometry (HR-CS-ET AAS), Molecular absorption spectrometry (MAS), Fluorine determination, Strontium monoflouride (SrF), Water sample, GRAPHITE-FURNACE, SPECTROMETRY, ALUMINUM, HALOGENS, FLAME
  • İstanbul Teknik Üniversitesi Adresli: Evet

Özet

The presence of fluorine (F) was detected via the rotational molecular absorption line of diatomic strontium-monofluoride (SrF) generated in the gas phase at 651.187 nm using high-resolution continuum source electrothermal atomic absorption spectrometry. Upon the addition of excess strontium (Sr) as the nitrate, the fluorine in the sample was converted to SrF in the gas phase of a graphite furnace. The effects on the accuracy, precision and sensitivity of variables such as the SrF wavelength, graphite furnace program, amount of Sr, coating of the graphite tube and platform with Zr and Ir and the use of a modifier were investigated and optimized. It was determined that there was no need to use a modifier or to cover the platform/tubes with Zr or Ir. Fluorine concentrations in various water samples (certified waste water, tap water, drinking water and mineral water) were determined using 20 mu g of Sr as the molecule-forming reagent and applying a maximum pyrolysis temperature of 800 degrees C and a molecule-forming temperature of 2200 degrees C with a heating rate of 2000 degrees C s(-1). Good linearity was maintained up to 0.1 mu g of F. The accuracy and precision of the method were tested by analyzing certified reference wastewater. The results were in good agreement with certified values, and the precision was satisfactory (RSD<10%). The limit of detection and the characteristic mass for the method were 0.36 ng and 0.55 ng, respectively. Finally, the fluorine concentrations in several drinking water and mineral water samples taken from the market were determined. The results were in good agreement with the values supplied by the producers. No significant differences were found between the results from the linear calibration and standard addition techniques. The method was determined to be simple, fast, accurate and sensitive. (C) 2012 Elsevier B.V. All rights reserved.