Direct determination of manganese in vitamin-mineral tablets using solid sampling electrothermal atomic absorption spectrometry


COSKUN N., Akman S.

TALANTA, vol.64, no.2, pp.496-500, 2004 (SCI-Expanded) identifier identifier

  • Publication Type: Article / Article
  • Volume: 64 Issue: 2
  • Publication Date: 2004
  • Doi Number: 10.1016/j.talanta.2004.03.012
  • Journal Name: TALANTA
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Page Numbers: pp.496-500
  • Istanbul Technical University Affiliated: Yes

Abstract

Manganese in vitamin-minerals tablets was determined by solid sampling electrothermal atomic absorption spectrometry (SS-ETAAS) using three different calibration methods, namely calibration against aqueous standards, standard addition with aqueous standards on solid samples and calibration against solid certified standards. Samples were only finely ground and introduced directly into the furnace by means of solid autosampler system without any dissolving process. Effects of different calibration techniques, temperatures and heating rates of atomization and pyrolysis steps on the accuracy and precision of the analyte elements were investigated. After optimization of the experimental parameters, there is good agreement (at 95% confidence level) between the results obtained by solid sampling and those obtained by acid digestion of samples. (C) 2004 Elsevier B.V. All rights reserved.