10(-12) level reproducibility of an iodine-stabilized He-Ne laser endorsed by absolute frequency measurements in the BIPM and UME


Hamid R., SABIN E., CELIK M., OZEN G., ZUCCO M., ROBERTSSON L., ...More

METROLOGIA, vol.43, no.1, pp.106-108, 2006 (SCI-Expanded) identifier identifier

  • Publication Type: Article / Article
  • Volume: 43 Issue: 1
  • Publication Date: 2006
  • Doi Number: 10.1088/0026-1394/43/1/015
  • Journal Name: METROLOGIA
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Page Numbers: pp.106-108
  • Istanbul Technical University Affiliated: Yes

Abstract

The absolute frequency of the f-component for the UME laser was measured with the BIPM femtosecond laser comb to be (473 612 353 602.0 +/- 1.1) kHz during 3 March to 8 March, 2003 and with the UME comb to be (473 612 353 600.6 +/- 1.1) kHz from 20 June to 25 June, 2005. After appropriate correction for power and modulation shifts a subkilohertz agreement between the two measurements was found. Moreover, the relative frequency stability for different UME-BIPM laser pairs is investigated using standard heterodyne techniques and Allan variance statistics.