International Conference on Advanced Mathematical and Computational Tool in Metrology and Testing, Paris, France, 01 June 2008, vol.78, pp.154-157
A procedure for optimal selection of sample input signals via D-optimality criterion to get the best calibration characteristics of measuring apparatus is proposed. As an example the problem of optimal selection of standard pressure setters when calibrating differential pressure gage is solved.