ELECTROANALYSIS, cilt.25, sa.7, ss.1796-1802, 2013 (SCI-Expanded)
Electrochemical reduction and determination of montelukast (MKS) was studied in methanol - 0.1M HCl solution (1:1, v/v) by cyclic voltammetry (CV) and differential pulse voltammetry (DPV) at chitosan modified carbon paste electrode. The linear range was 1.70x10(-7)-1.83x10(-5)M for DPV analysis. Limit of detection (LOD) and limit of quantification (LOQ) were 5.32x10(-8)M and 1.61x10(-7)M, respectively. The developed method was successfully applied to the determination of MKS in tablets and spiked human plasma. The results obtained were in good agreement with those obtained using a reported spectrofluorimetric technique.